Particle size distribution analysis showing D10, D50 and D90 measurements at SGM National Laboratory.

 D10, D50 and D90 are the sizes below which 10%, 50% and 90% of the particles (by volume) fall. D50 is the median size, D10 shows the fine end and D90 the coarse end. Span = (D90 − D10) / D50 tells you how wide the distribution is.

Let’s be honest. Most people open a particle size report, skip the graph, and hunt for one number they recognise. Usually that’s D50. Then they wonder what the other two are for. They’re there for a reason. In particle size distribution analysis, D10, D50 and D90 together tell you far more than any single value can. Below, we explain each one, run a worked example, and answer the questions we hear most from people using our particle size analysis services.

What Is Particle Size Distribution Analysis?

Here’s the thing about powders: no two particles are the same size. Grind calcium carbonate as carefully as you like and you’ll still get fine dust, a big middle crowd and a few stubborn coarse grains.

Particle size distribution analysis measures that whole mix and gives you a curve. D10, D50 and D90 are three points picked off it, so buyers, suppliers and QC teams can talk about the same powder without passing graphs around.

How We Actually Measure It

Mostly, it’s laser diffraction. We disperse the sample in water, alcohol or air, shine a laser through it and record the scattered light. Small particles scatter at wide angles, big ones at narrow angles, and the software works backwards to a size distribution.

Sieving and image analysis have their place, but for day-to-day particle size distribution analysis, laser diffraction wins on speed and repeatability. Any well-run particle size analysis lab in India should be following ISO 13320 for it.

What Do D10, D50 and D90 Mean?

Picture every particle lined up from smallest to largest. Walk along the line adding up volume. At 10% of the total, you’ve hit D10. Halfway is D50. At 90%, you’ve found D90.

D10: The Fine End

D10 is the size below which 10% of your material sits. So if your D10 is 2 µm, a tenth of the powder is finer than 2 µm. If you’re fighting dust, caking or poor flow, look here first. Fines are nearly always behind it.

D50: The Median

D50 cuts the sample in half. Half the volume is finer, half is coarser. When someone says “it’s a 10-micron powder”, this is usually what they mean.

D90: The Coarse End

D90 is the size below which 90% of the material falls, which means only the top 10% is bigger. In particle size distribution analysis, it’s your early warning for grit and oversize particles.

You may see Dv10, Dv50 and Dv90 instead. The “v” just means volume-weighted, the normal output when particle size distribution analysis is done by laser diffraction.

A Worked Example With Real Numbers

Numbers make this easier. Here are two batches of ground calcium carbonate filler, tested by laser diffraction using the same wet method.

Particle Size Distribution Comparison
Particle Size Distribution Comparison
Parameter Batch A Batch B
D10 2.1 µm 4.0 µm
D50 8.4 µm 8.2 µm
D90 24.6 µm 14.5 µm
Span 2.68 1.28
D[4,3] (Volume Mean) 11.3 µm 8.9 µm

If you only checked D50, you’d say these two were the same powder. They’re not. Batch A has more fines at the bottom and a long coarse tail at the top, with D90 close to 25 µm. In a paint or plastic, that could mean grit and lower gloss. Same “size” on paper, very different behaviour in the plant.

That’s why particle size distribution analysis reports three values, and why good particle size analysis services never hand you just one.

Reading an Annotated Malvern Report

A Malvern Mastersizer report from particle size distribution analysis looks busy but follows the same layout every time. Ask your particle size analysis lab in India for the full PDF rather than a one-line summary, then look for:

  1. Sample and method details: dispersant, refractive index, obscuration and stirrer speed. If these aren’t recorded, the result can’t be repeated.
  2. The results table: Dx(10), Dx(50) and Dx(90), along with D[4,3], D[3,2] and span.
  3. The frequency curve: one smooth hump means one population. Two humps usually mean agglomerates or a blend.
  4. The cumulative curve: read across at 10%, 50% and 90%, drop down to the size axis, and you’ve found D10, D50 and D90 yourself.
  5. Quality indicators such as weighted residual. The lower it is, the better the fit.

How Do I Calculate Span From D10, D50 and D90?

Span tells you how spread out your powder is. It’s the quickest width check in particle size distribution analysis, and most particle size analysis services print it anyway. Here’s the formula:

Span = (D90 − D10) / D50

Batch A: 24.6 − 2.1 = 22.5. Divide by 8.4 and you get 2.68.

Batch B: 14.5 − 4.0 = 10.5. Divide by 8.2 and you get 1.28.

Roughly, under 1 is narrow, 1 to 2 is moderate, and over 2 is broad. None is “good” or “bad” by itself. Broad spreads pack tightly, which suits ceramics and concrete. Narrow ones behave predictably, which pharma and pigment makers prefer.

Is D50 the Same as Average Particle Size?

No, and honestly, this one causes more confusion than anything else in particle size distribution analysis. D50 is the median. The “average” on a laser diffraction report is normally D[4,3], the volume-weighted mean.

Look at Batch A. D50 is 8.4 µm, but D[4,3] is 11.3 µm. The coarse tail drags the mean up while the median hardly moves. So if D[4,3] sits well above D50, something coarse is hiding in there.

There’s also D[3,2], the Sauter mean, which leans towards fines and matters when surface area does. Whenever you order particle size analysis services or write a spec, say exactly which value you mean. And ask your particle size analysis lab in India which mean it reports. “Average size 10 µm” with no definition has started more supplier arguments than we can count.

Why Is D90 Important in Quality Control?

Because that’s where trouble lives. In particle size distribution analysis for QC, D90 is often the most watched number. Complaints rarely come from the median; they come from the coarse tail. For example:

  • Paints and coatings: oversize particles show up as specks and failed fineness-of-grind checks.
  • Pharmaceuticals: coarse API particles dissolve slowly and can upset content uniformity. That’s why so many drug specs carry a D90 limit.
  • Fire extinguisher dry powders: coarse lumps hurt flow and discharge through the nozzle.
  • Filtration and spraying: big particles block filters and spray tips.

Watching D90 batch to batch is also a quiet early warning. A slow creep upward often means mill wear or a torn sieve, and you catch it before your customer does. A dependable particle size analysis lab in India can track that trend for you across regular production samples

What Is a Good D50 for My Powder?

We get asked this a lot. Honestly, it depends on what the powder has to do. The D50 you aim for in particle size distribution analysis is set by the end use. Typical ranges:

  • Fine fillers for paints and plastics: often 1 to 10 µm.
  • Cement: commonly around 10 to 20 µm.
  • Inhalation drug particles: usually 1 to 5 µm, so they actually reach the lungs.
  • Coarser mineral grades for construction or foundry use: anywhere from tens to hundreds of microns.

Start from the application and your customer’s spec, ideally with help from your particle size analysis lab in India. Pick a D50 target, then set limits on D10 and D90. Not sure where to start? Have a particle size analysis lab in India test your best batch and your worst batch side by side. The difference usually tells you where the line belongs.

Choosing Particle Size Analysis Services

A result from particle size distribution analysis is only as good as the method behind it. Two labs can report different D50 values on the same sample just because of a different dispersant or refractive index. So when you compare particle size analysis services, ask:

  • Is the lab NABL accredited for this test?
  • Do they follow ISO 13320 and develop a dispersion method for your material?
  • Will they share the full particle size distribution analysis report, with the curve and quality indicators?
  • Can they repeat the measurement and show you it’s reproducible?

At SGM National Lab, our particle size analysis lab in India runs laser diffraction on minerals, chemicals, pharmaceuticals and industrial powders, and every report carries its method details. Whether it’s a one-off check or routine batch testing, we’ll help you set D10, D50 and D90 limits that make sense. If you’d like a particle size analysis lab in India that explains your results instead of just printing them, and offers particle size analysis services you can rely on batch after batch, we’re always happy to go through a report with you.

Conclusion

Particle size distribution analysis takes a messy curve and boils it down to three numbers you can actually work with. D10 shows you the fines, D50 gives you the middle, and D90 warns you about the coarse tail. Add span and you know how wide the spread is. Together, they show why two powders with the same “average” can behave so differently.

Good particle size distribution analysis comes down to the right method, a complete report and limits that match how your product is used. That’s what you should expect from professional particle size analysis services. Pick a particle size analysis lab in India that’s open about its method, keep up regular particle size analysis services for your production batches, and your D10, D50 and D90 values will mean the same thing every single time.

FAQs

It measures how much of a powder or suspension falls into each size range, giving a distribution instead of a single average number.

D50 is the median size, so half the particles are smaller. D10 and D90 mark the sizes below which 10% and 90% of particles fall, describing the fine and coarse ends.

Size affects flow, dissolution, reactivity, packing, colour and strength, so it can change product quality in minerals, fertilizers, paints and ceramics.

Common methods are sieve analysis, laser diffraction, sedimentation and microscopy. The right one depends on the size range and the material.

Span = (D90 - D10) / D50 and shows how wide the distribution is. A lower span means a narrower, more uniform powder. The ideal value depends on your application.

Yes. Dispersion in liquid can break up agglomerates that stay clumped in dry mode, so choose the method that matches how the material is used.

Mention the material type, expected size range, whether it dissolves in water and the report format you need, such as D10/D50/D90 or a full curve.